    Rheometer Installation Requirements
                                    
   
 
  

1. House air rated at 90 psig (6 bar)

  • NOVA/DYNALYSER/VISCOANALYSER
  • STRESSTECH/VISCOTECH
  • DSR QC

(*Note: Depending on ambient conditions, tap water may be used for faster cooling.)



2.5 CFM
2.5 CFM
1.0 CFM*

2. 110V 15A electrical circuit with a minimum of 6 receptacles (or 2 receptacles and a power strip).

(*Note: A dedicated, isolated circuit and circuit breaker with variation in line voltage of less that 1% is required in order to avoid inadequate power situations.

3. Five (5) feet of usable bench space with access below bench top for temperature control circulator (if ordered) and drive motor cooling system.

4. Pentium computer or equivalent, a minimum of 256MB ram (512MB ram is required for the Fast Oscillation test mode), 3.5" drive, SVGA monitor, mouse, modem, 1 open com port (9 pin sub-D), and Windows 2000, or XP operating systems.

5. Water and Oil filters for Air line
Manufacturer Hankison International, Tel: 724 745 1555
    A. 1.0 micron water filter (P/N HF7-12-3-DPL)
    B. 0.025 micron oil filter (P/N HF5-16-3-DPL)

Notes 1: Filter (B) is downstream of filter (A). ATS RheoSystems requires a 3/8 NPT female fitting (standard on filter B) for connection to the Rheometer. Air filters must be installed prior to arrival date of instrument installation.

6. For NOVA/DYNALYSER/VISCOANALYSER/VISCOTECH unit, two (2) ethernet connection ports in the Windows computer if the computer used for the instrument will be connected to a Network. One connection will be used for the Network and the other will be used for the Rheometer unit. The connection for the Rheometer will use a fixed IP address. This address is: 192.168.0.150

7. For STRESSTECH/DSR QC, 1 open com port (9 pin sub –D) is required.


Please acknowledge acceptance of these installation requirements by signing and returning this document to ATS RheoSystems/REOLOGICA Instruments. Receipt of this signed acknowledgement is required prior to scheduling of installation.